Publicação: A novel device for greater precision and safety in open-wedge high tibial osteotomy: cadaveric study
dc.contributor.author | Ribeiro, Cristiano Hossri | |
dc.contributor.author | Mod, Mauricio Sante Bettio | |
dc.contributor.author | Isch, Daniel | |
dc.contributor.author | Baier, Clemens | |
dc.contributor.author | Maderbacher, Guenther | |
dc.contributor.author | Severino, Nilson Roberto | |
dc.contributor.author | Cataneo, Daniele Cristina [UNESP] | |
dc.contributor.institution | Hospital Orto | |
dc.contributor.institution | Ápice Clinic | |
dc.contributor.institution | Safrima | |
dc.contributor.institution | Asklepios Klinik | |
dc.contributor.institution | Santa Casa de Misericórdia de São Paulo (SCSP) | |
dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
dc.date.accessioned | 2020-12-12T02:23:44Z | |
dc.date.available | 2020-12-12T02:23:44Z | |
dc.date.issued | 2020-02-01 | |
dc.description.abstract | Introduction: The complications of the open-wedge osteotomy technique (slope, hypo- and hypercorrection and fractures) are related to difficulties in the control of the open-wedge angle during surgery. Materials and methods: In this cadaveric study, we evaluated the safety and precision of a novel system, the Realignment High Control System (RHC), in the correction of knee mechanical axis and slope. The RHC has a fixation plate coupled to a dynamic device that opens the osteotomy continuously, allowing plate fixation before osteotomy wedge opening. Results: All procedures were easily performed, with no fractures. The openings equaled the indicated by the navigation system for 7.5° and 10°, and for 5°, there was a statistically, although not clinically significant, difference of 0.6°. The slope shown by the RHC setting and in the navigation system was significantly different for the 10° setting only, with a mean difference of 0.563°. Conclusions: RHC facilitates the surgical technique of high tibial osteotomy, with gradual wedge opening, precise correction of the mechanical axis, and appropriate control of the tibial slope, even with larger openings. | en |
dc.description.affiliation | Knee Surgery Division Hospital Orto, Av. Nove de Julho, 1017, Vila Icaraí | |
dc.description.affiliation | Ápice Clinic | |
dc.description.affiliation | Safrima | |
dc.description.affiliation | Asklepios Klinik | |
dc.description.affiliation | Department of Orthopedics and Traumatology Santa Casa de Misericórdia de São Paulo (SCSP) | |
dc.description.affiliation | Surgery and Orthopaedics Department Universidade Estadual de São Paulo (Unesp) | |
dc.description.affiliationUnesp | Surgery and Orthopaedics Department Universidade Estadual de São Paulo (Unesp) | |
dc.format.extent | 203-208 | |
dc.identifier | http://dx.doi.org/10.1007/s00402-019-03300-w | |
dc.identifier.citation | Archives of Orthopaedic and Trauma Surgery, v. 140, n. 2, p. 203-208, 2020. | |
dc.identifier.doi | 10.1007/s00402-019-03300-w | |
dc.identifier.issn | 1434-3916 | |
dc.identifier.issn | 0936-8051 | |
dc.identifier.scopus | 2-s2.0-85075161225 | |
dc.identifier.uri | http://hdl.handle.net/11449/201086 | |
dc.language.iso | eng | |
dc.relation.ispartof | Archives of Orthopaedic and Trauma Surgery | |
dc.source | Scopus | |
dc.subject | Computer-assisted surgery | |
dc.subject | Genu varum | |
dc.subject | High-tibial-osteotomy | |
dc.subject | Knee osteoarthritis | |
dc.subject | Surgical technique | |
dc.title | A novel device for greater precision and safety in open-wedge high tibial osteotomy: cadaveric study | en |
dc.type | Artigo | |
dspace.entity.type | Publication | |
unesp.author.orcid | 0000-0003-4021-1198[1] | |
unesp.campus | Universidade Estadual Paulista (Unesp), Faculdade de Medicina, Botucatu | pt |
unesp.department | Cirurgia e Ortopedia - FMB | pt |