Determination of iodide and idoxuridine at a glutaraldehyde-cross-linked poly-L-lysine modified glassy carbon electrode

Nenhuma Miniatura disponível

Data

2005-07-01

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Wiley-Blackwell

Tipo

Artigo

Direito de acesso

Acesso restrito

Resumo

The detection limit (about 0.017 mu g mL(-1)) for voltammetric determination of iodide (peak at +0.87 V vs. Ag/AgCl at pH 2) at a glutaraldehyde-cross-linked poly-L-lysine modified glassy carbon electrode involving oxidation to iodine was found to be several orders of magnitude lower than that for the voltammetric determination on a bare glassy carbon electrode. This method was applied successfully to the determination of iodide in two medicinal formulations. Idoxuridine was determined indirectly at the same electrode by accumulating it first at -0.8 V vs. Ag/AgCl. At this potential the C-I bond in the adsorbed idoxuridine is reduced giving iodide, which is then determined at the modified electrode. The method was successfully applied to the determination of idoxuridine in a urine sample.

Descrição

Idioma

Inglês

Como citar

Electroanalysis. Weinheim: Wiley-v C H Verlag Gmbh, v. 17, n. 14, p. 1309-1316, 2005.

Itens relacionados

Financiadores