Repository logo

Use of carbon dioxide (CO2) laser radiation in exposure of implants (stage II): A metallography evaluation in dogs

Loading...
Thumbnail Image

Advisor

Coadvisor

Graduate program

Undergraduate course

Journal Title

Journal ISSN

Volume Title

Publisher

Laser Inst Amer

Type

Article

Access right

Acesso restrito

Abstract

In this study we analyzed possible damages that vaporization from laser radiation could cause to implant material. Fifteen standard titanium implants, measuring 3.75 mm in diameter by 7 mm in length, were placed into the upper and lower jaws of three dogs according to Branemark's system. After osseointegration, all implants were exposed. In group I (control) conventional exposure with a punch was used; in group II, a CO2 laser with 2 W (power density: 256 W/cm(2); fluency: 0.077 J/cm(2), and a pulse mode of 0.30 ms) was used, and in group III 4 W (power density: 512 W/cm(2), fluency: 0.154 J/cm(2), and a pulse mode of 0.30 ms) was used. After vaporization, the cover screws were removed and sent for metallographic examination. The results showed that cover screws irradiated with 2 and 4 W power caused no superficial or microstructural alteration. The results also showed that the prescribed power densities, fluencies, and the use of the pulse mode were suitable for exposing implants without damage to tissue or implant material. (C) 2002 Laser Institute of America.

Description

Keywords

osseointegrated implants, carbon dioxide laser, stage II exposed implants

Language

English

Citation

Journal of Laser Applications. Orlando: Laser Inst Amer, v. 14, n. 4, p. 248-251, 2002.

Related itens

Sponsors

Units

Departments

Undergraduate courses

Graduate programs

Other forms of access