Avaliação da resistência ao arrancamento de mini-implantes ortodônticos: estudo laboratorial

Loading...
Thumbnail Image

Advisor

Santos, Eduardo Cesar Almada
Santos, Paulo Henrique dos

Coadvisor

Graduate program

Odontologia - FOA

Undergraduate course

Journal Title

Journal ISSN

Volume Title

Publisher

Universidade Estadual Paulista (Unesp)

Type

Master's thesis

Access right

Acesso abertoAcesso Aberto

Abstract

Abstract (portuguese)

Objetivo: comparar, por meio do teste de arrancamento, a estabilidade primária de mini-implantes com diâmetro e comprimento diferentes inseridos em osso artificial. Material e método: foram utilizados 80 mini-implantes da marca SIN (Sistema de Implante) com 1mm de perfil transmucoso e divididos em 4 grupos: grupo I - mini-implantes de 1,4 mm de diâmetro e 6 mm de comprimento ; o grupo II - mini-implantes de 1,4 mm de diâmetro e 8 mm de comprimento; o grupo III - mini-implantes de 1,6 mm de diâmetro e 6 mm de comprimento; grupo IV - mini-implantes de 1,6 mm de diâmetro e 8 mm de comprimento. Os mini-implantes foram inseridos em bloco de osso artificial Sawbones (SAWBONES Pacific Research Laboratories) e submetidos ao teste de arrancamento na máquina de ensaio universal – EMIC. Resultados: os valores médios de força de arrancamento dos mini-implantes oscilaram entre 621.995N a 738.220N. A Análise de Variância mostrou que houve diferença estatisticamente significante entre os grupos (p<0.05). Na comparação entre os grupos houve igualdade entre os grupos I e III. Todas as demais comparações foram estatisticamente diferentes. Conclusões: em relação à estabilidade primária, constatou-se que os grupos I e III foram iguais, enquanto que, todos os demais apresentaram diferença na estabilidade

Abstract (english)

Objective: to compare, by means of the pullout test, the primary stability of mini-implants with different diameter and length placed in artificial bone. Method: 80 mini-implants Sin (Implant System) with 1mm transmucosal profile and divided into four groups: Group I - mini-implants with diameter of 1.4mm and 6 mm of long; Group II - mini-implants with diameter of 1.4mm and 8 mm long; Group III - mini-implants with diameter of 1.6mm and 6 mm of long; Group IV - mini-implants with a diameter of 1.6mm and 8 mm of long. Mini-implants were inserted in block artificial bone Sawbones (Sawbones Pacific Research Laboratories) and tested in pullout universal testing machine – EMIC. Results: the mean pullout strength of mini-implants ranged from 621.995 to 738,220 N. Analysis of Variance showed that there was a statistically significant difference between the groups (p <0.05). Comparing the groups were only equality between groups I (mini-implants 1.4 Ø x 6 mm) and III (mini-implants Ø 1.6 x 6mm). All other comparisons are statistically different. Conclusions: relative primary stability, it was found that the groups I and III were similar, whereas all others show differences in stability

Description

Keywords

Procedimentos de ancoragem ortodôntica, Implantes dentários, Ortodontia, Orthodontic anchorage procedures, Dental implantation

Language

Portuguese

Citation

GOUVÊA, Carla Corrêa Mendes. Avaliação da resistência ao arrancamento de mini-implantes ortodônticos: estudo laboratorial. 2011. 54 f. Dissertação (mestrado) - Universidade Estadual Paulista, Faculdade de Odontologia de Araçatuba, 2011.