Publicação:
A MICROSTRAIN COMPARISON OF PASSIVELY FITTING SCREW-RETAINED AND CEMENTED TITANIUM FRAMEWORKS

dc.contributor.authorVasconcellos, Diego Klee de [UNESP]
dc.contributor.authorKojima, Alberto Noriyuki [UNESP]
dc.contributor.authorMelo Mesquita, Alfredo Mikail
dc.contributor.authorBottino, Marco Antonio [UNESP]
dc.contributor.authorOezcan, Mutlu
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionPaulista Univ UNIP
dc.contributor.institutionUniv Zurich
dc.date.accessioned2015-03-18T15:54:01Z
dc.date.available2015-03-18T15:54:01Z
dc.date.issued2014-10-01
dc.description.abstractStatement of problem. An imprecise fit between frameworks and supporting dental implants in loaded protocols increases the strain transferred to the periimplant bone, which may impair healing or generate microgaps.Purpose. The purpose of this study was to investigate the microstrain between premachined 1-piece screw-retained frameworks (group STF) and screw-retained frameworks fabricated by cementing titanium cylinders to the prefabricated framework (group CTF). This procedure was developed to correct the misfit between frameworks and loaded implants.Material and methods. Four internal hexagon cylindrical implants were placed 10 mm apart in a polyurethane block by using the surgical guides of the corresponding implant system. Previously fabricated titanium frameworks (n=10) were divided into 2 groups. In group STF, prefabricated machined frameworks were used (n=5), and, in group CTF, the frameworks were fabricated by using a passive fit procedure, which was developed to correct the misfit between the cast titanium frameworks and supporting dental implants (n=5). Both groups were screw-retained under torque control (10 Ncm). Six strain gauges were placed on the upper surface of the polyurethane block, and 3 strain measurements were recorded for each framework. Data were analyzed with the Student t test (alpha=.05).Results. The mean microstrain values between the framework and the implants were significantly higher for group STF (2517 me) than for group CTF (844 me) (P<.05).Conclusions. Complete-arch implant frameworks designed for load application and fabricated by using the passive fit procedure decreased the strain between the frameworks and implants more than 1 piece prefabricated machined frameworks.en
dc.description.affiliationSao Paulo State Univ, Dent Sch UNESP, Dept Dent Mat & Prosthodont, Sao Jose Dos Campos, Brazil
dc.description.affiliationPaulista Univ UNIP, Dept Dent Mat & Prosthodont, Sao Paulo, Brazil
dc.description.affiliationUniv Zurich, Clin Fixed & Removable Prosthodont & Dent Mat Sci, Ctr Dent & Oral Med, Dent Mat Unit, Zurich, Switzerland
dc.description.affiliationUnespSao Paulo State Univ, Dent Sch UNESP, Dept Dent Mat & Prosthodont, Sao Jose Dos Campos, Brazil
dc.format.extent834-838
dc.identifierhttp://dx.doi.org/10.1016/j.prosdent.2013.12.018
dc.identifier.citationJournal Of Prosthetic Dentistry. New York: Mosby-elsevier, v. 112, n. 4, p. 834-838, 2014.
dc.identifier.doi10.1016/j.prosdent.2013.12.018
dc.identifier.issn0022-3913
dc.identifier.lattes9234456003563666
dc.identifier.urihttp://hdl.handle.net/11449/116722
dc.identifier.wosWOS:000342543500024
dc.language.isoeng
dc.publisherElsevier B.V.
dc.relation.ispartofJournal Of Prosthetic Dentistry
dc.relation.ispartofjcr2.347
dc.relation.ispartofsjr1,087
dc.rights.accessRightsAcesso restrito
dc.sourceWeb of Science
dc.titleA MICROSTRAIN COMPARISON OF PASSIVELY FITTING SCREW-RETAINED AND CEMENTED TITANIUM FRAMEWORKSen
dc.typeArtigo
dcterms.licensehttp://www.elsevier.com/about/open-access/open-access-policies/article-posting-policy
dcterms.rightsHolderElsevier B.V.
dspace.entity.typePublication
unesp.author.lattes9234456003563666
unesp.author.lattes6168075208730450[2]
unesp.author.orcid0000-0003-4607-9095[2]
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Ciência e Tecnologia, São José dos Campospt
unesp.departmentMateriais Odontológicos e Prótese - ICTpt