Logotipo do repositório
 

Publicação:
Effect of the number of screws on the stability of locking mandibular reconstruction plates

dc.contributor.authorPereira-Filho, V. A. [UNESP]
dc.contributor.authorDa Silva, B. N. [UNESP]
dc.contributor.authorNunes Reis, J. M. [UNESP]
dc.contributor.authorSpin-Neto, R. [UNESP]
dc.contributor.authorReal Gabrielli, M. F. [UNESP]
dc.contributor.authorMonnazzi, M. S. [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2014-05-27T11:28:42Z
dc.date.available2014-05-27T11:28:42Z
dc.date.issued2013-03-25
dc.description.abstractAtrophic mandible fractures are frequently a challenge to stabilize. This study evaluated, through mechanical testing in vitro, the number of locking screws that is sufficient to withstand loading when applied with a locking reconstruction plate in the fixation of atrophic mandible fractures. Polyurethane mandibles with a simulated linear fracture at the midline were used as substratum. Results show that resistance of the fixation is poor when one and two screws are used on each side of the fracture. Three screws on each side of the fracture significantly increases the resistance to displacement. However, no additional strength is added to the construct when more than three screws per side are used. © 2013 International Association of Oral and Maxillofacial Surgeons.en
dc.description.affiliationOral and Maxillofacial Surgery Department Dental School of Araraquara (UNESP), Rua Voluntarios da Patria, 2777, Ap 1001, Araraquara, São Paulo
dc.description.affiliationUnespOral and Maxillofacial Surgery Department Dental School of Araraquara (UNESP), Rua Voluntarios da Patria, 2777, Ap 1001, Araraquara, São Paulo
dc.format.extent732-735
dc.identifierhttp://dx.doi.org/10.1016/j.ijom.2013.02.010
dc.identifier.citationInternational Journal of Oral and Maxillofacial Surgery, v. 42, n. 6, p. 732-735, 2013.
dc.identifier.doi10.1016/j.ijom.2013.02.010
dc.identifier.issn0901-5027
dc.identifier.issn1399-0020
dc.identifier.lattes8029177169916525
dc.identifier.scopus2-s2.0-84877017709
dc.identifier.urihttp://hdl.handle.net/11449/74867
dc.identifier.wosWOS:000319538000009
dc.language.isoeng
dc.relation.ispartofInternational Journal of Oral and Maxillofacial Surgery
dc.relation.ispartofjcr2.164
dc.relation.ispartofsjr1,137
dc.rights.accessRightsAcesso restrito
dc.sourceScopus
dc.subjectatrophic mandible
dc.subjectedentulous mandible
dc.subjectmandibular fracture
dc.subjectpolyurethan
dc.subjectbone plate
dc.subjectcontrolled study
dc.subjectfracture fixation
dc.subjecthuman
dc.subjectin vitro study
dc.subjectlocking screw
dc.subjectmandible fracture
dc.subjectmandible reconstruction
dc.titleEffect of the number of screws on the stability of locking mandibular reconstruction platesen
dc.typeArtigo
dcterms.licensehttp://www.elsevier.com/about/open-access/open-access-policies/article-posting-policy
dspace.entity.typePublication
unesp.author.lattes8029177169916525
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Odontologia, Araraquarapt
unesp.departmentDiagnóstico e Cirurgia - FOARpt