A complete dosimetric characterization of two Sr-90-Y-90 dermatologic applicators
Loading...
Files
External sources
External sources
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier B.V.
Type
Article
Access right
Acesso restrito
Files
External sources
External sources
Abstract
A complete dosimetric characterization of two Amershan Sr-90-Y-90 dermatologic applicators is described in this present work. The dosimetric parameters analyzed are: percentage depth dose curve, radial dose distribution, non-uniformity and asymmetry. Both applicators are planar-circular having 22.57 and 9.0 mm diameters. In the range where the percentage depth dose goes from 100% down to 20%, the measured percentage depth dose and that obtained by the Monte Carlo simulation have shown maximum discrepancy of 5.3% for both applicators. The radial dose distribution has been measured at several depths using a GafChromic (R) EBT QD+ films and it was also calculated by simulation. The discrepancies found did not exceed 5.9% up to the depth of 1.8 mm, where the percentage depth dose drops to 40% of the maximum. The maximum non-uniformity and asymmetry are 1.7% and 5.3% for the first applicator and 22.7% and 25.9% for the second applicator, respectively. Both applicators meet the specification for the maximum non-uniformity established by the adopted protocol, whose limit is 30%. As for the asymmetry the limit is 20% and the second applicator exceeded it in about 5.9%. (C) 2010 Elsevier B.V. All rights reserved.
Description
Keywords
Betatherapy, Monte Carlo method, Keloids, Pterygium, Dosimetry
Language
English
Citation
Nuclear Instruments & Methods In Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment. Amsterdam: Elsevier B.V., v. 632, n. 1, p. 147-151, 2011.