Logo do repositório

Can the modeling for simplification of a dental implant surface affect the accuracy of 3D finite element analysis?

Resumo

The aim of this study was to assess stress/strain of different implant modeling simplifications by 3D-FEA. Three variation of external hexagon implant (à 3.75 à 10 mm) supporting one molar crown were simulated: A (no threads); B (slightly threads simplification); C (original design). 200 N (axial) and 100 N (oblique) were applied. Cortical bone was evaluated by maximum principal stress and microstrain qualitatively and quantitatively (ANOVA and Tukey post hoc (p < 0.05)). Higher stress levels (p < 0.05) were observed in model A. Models B and C presented similar stress transmission. It was possible to conclude that slightly simplification should be used for studies evaluating stress transferring for bone tissue.

Descrição

Palavras-chave

biomechanics, Dental implants, finite element analysis

Idioma

Inglês

Citação

Computer Methods in Biomechanics and Biomedical Engineering, v. 19, n. 15, p. 1665-1672, 2016.

Itens relacionados

Financiadores

Unidades

Item type:Unidade,
Faculdade de Odontologia
FOA
Campus: Araçatuba


Cursos de graduação

Programas de pós-graduação