Can the modeling for simplification of a dental implant surface affect the accuracy of 3D finite element analysis?
Carregando...
Fontes externas
Fontes externas
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Taylor & Francis
Tipo
Artigo
Direito de acesso
Acesso aberto

Fontes externas
Fontes externas
Resumo
The aim of this study was to assess stress/strain of different implant modeling simplifications by 3D-FEA. Three variation of external hexagon implant (à 3.75 à 10 mm) supporting one molar crown were simulated: A (no threads); B (slightly threads simplification); C (original design). 200 N (axial) and 100 N (oblique) were applied. Cortical bone was evaluated by maximum principal stress and microstrain qualitatively and quantitatively (ANOVA and Tukey post hoc (p < 0.05)). Higher stress levels (p < 0.05) were observed in model A. Models B and C presented similar stress transmission. It was possible to conclude that slightly simplification should be used for studies evaluating stress transferring for bone tissue.