Repository logo

Development of a voltammetric sensor for chromium(VI) determination in wastewater sample

Loading...
Thumbnail Image

Advisor

Coadvisor

Graduate program

Undergraduate course

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier B.V.

Type

Article

Access right

Acesso restrito

Abstract

Screen-printed carbon electrode (SPCE) modified with poly-L-histidine film can be successfully applied for chromium(VI) determination based on its pre-concentration. Optimum adherence and stability of the POIY-L-histidine film was obtained by direct addition of PH solution 1% (w/v) on the electrode surface, followed by heating at 80 degrees C during 5 min. Linear response range, sensitivity and limit of detection were 0. 1-150 [mu mol L-1, 4. 13 LA mu mol L` and 0.046 mu mol L-1. The repeatability of the proposed sensor, evaluated in terms of relative standard deviation, was measured as 3.2% for 10 experiments in 40 mu mol L-1 using the same electrode and 4.0% using screen-printed electrode as disposable sensor, respectively. The voltammetric sensor was applied to determination of Cr(VI) and indirect determination of Cr(III) in wastewater samples previously treated by a leather dyeing industry and the average recovery for these samples was around 97%. (C) 2006 Elsevier B.V. All rights reserved.

Description

Keywords

chromium(VI), stripping analysis, screen-printed modified electrode, poly-L-histidine

Language

English

Citation

Sensors and Actuators B-chemical. Lausanne: Elsevier B.V. Sa, v. 123, n. 2, p. 902-908, 2007.

Related itens

Sponsors

Units

Item type:Unit,
Instituto de Química
IQAR
Campus: Araraquara


Departments

Undergraduate courses

Graduate programs

Other forms of access