Development of a voltammetric sensor for chromium(VI) determination in wastewater sample
Loading...
Files
External sources
External sources
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier B.V.
Type
Article
Access right
Acesso restrito
Files
External sources
External sources
Abstract
Screen-printed carbon electrode (SPCE) modified with poly-L-histidine film can be successfully applied for chromium(VI) determination based on its pre-concentration. Optimum adherence and stability of the POIY-L-histidine film was obtained by direct addition of PH solution 1% (w/v) on the electrode surface, followed by heating at 80 degrees C during 5 min. Linear response range, sensitivity and limit of detection were 0. 1-150 [mu mol L-1, 4. 13 LA mu mol L` and 0.046 mu mol L-1. The repeatability of the proposed sensor, evaluated in terms of relative standard deviation, was measured as 3.2% for 10 experiments in 40 mu mol L-1 using the same electrode and 4.0% using screen-printed electrode as disposable sensor, respectively. The voltammetric sensor was applied to determination of Cr(VI) and indirect determination of Cr(III) in wastewater samples previously treated by a leather dyeing industry and the average recovery for these samples was around 97%. (C) 2006 Elsevier B.V. All rights reserved.
Description
Keywords
chromium(VI), stripping analysis, screen-printed modified electrode, poly-L-histidine
Language
English
Citation
Sensors and Actuators B-chemical. Lausanne: Elsevier B.V. Sa, v. 123, n. 2, p. 902-908, 2007.