Logotipo do repositório
 

Publicação:
Influence of Three Types of Abutments on Preload Values Before and After Cyclic Loading with Structural Analysis by Scanning Electron Microscopy

dc.contributor.authorButignon, Luis Eduardo [UNESP]
dc.contributor.authorBasilio, Mariana de Almeida [UNESP]
dc.contributor.authorPereira, Rodrigo de Paula [UNESP]
dc.contributor.authorArioli Filho, João Neudenir [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2014-12-03T13:10:49Z
dc.date.available2014-12-03T13:10:49Z
dc.date.issued2013-05-01
dc.description.abstractPurpose: To evaluate the effectiveness of three types of abutments in the maintenance of screw joint preload before and after cyclic loading as well as to observe possible microdamage in the structure of the components using a scanning electron microscope (SEM). Materials and Methods: Forty-five external-hex implants were embedded in epoxy resin, received their respective abutments, and were randomly divided into three experimental groups (n = 15): (1) machined titanium (Ti) abutments; (2) pre-machined gold (Au) abutments; and (3) machined zirconia (ZrO2) abutments. The abutment screws were tightened according to the manufacturer's recommended torque. Initially, a static bending test was performed using five specimens of each group to determine the load applied in the cyclic loading test. Thus, 10 specimens of each group were used to measure the reverse torque value (preload) of the abutment screw before and after loading. A cyclic loading (0.5 x 10(6) cycles; 15 Hz) between 11 and 211 N was applied at an angle of 30 degrees to the long axis of the implants. The group means were compared using analysis of variance and the Tukey test (alpha = .05). Results : The reverse torque analysis before cyclic loading showed no significant difference among the groups (P > .05). After cyclic loading, all preload means decreased significantly. The lowest decrease in preload was observed in the Ti group, whereas the highest decrease was observed in the ZrO2 group, with a significant difference noted between them (P = .010). The Au group presented an intermediate decrease, with no significant difference compared to the other groups (P > .05). SEM images showed structural changes in the mating surfaces of the abutments after cyclic loading. Conclusion: The load application reduced the preload means significantly in all groups, and more significantly in the ZrO2 group.en
dc.description.affiliationUniv Estadual Paulista, UNESP, Araraquara Dent Sch, Dept Dent Mat & Prosthodont, Sao Paulo, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, UNESP, Araraquara Dent Sch, Dept Dent Mat & Prosthodont, Sao Paulo, Brazil
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
dc.description.sponsorshipIdFAPESP: 09/03421-1
dc.format.extentE161-E170
dc.identifierhttp://dx.doi.org/10.11607/jomi.2481
dc.identifier.citationInternational Journal Of Oral & Maxillofacial Implants. Hanover Park: Quintessence Publishing Co Inc, v. 28, n. 3, p. E161-E170, 2013.
dc.identifier.doi10.11607/jomi.2481
dc.identifier.issn0882-2786
dc.identifier.lattes6380225075834021
dc.identifier.urihttp://hdl.handle.net/11449/112545
dc.identifier.wosWOS:000322078500007
dc.language.isoeng
dc.publisherQuintessence Publishing Co Inc
dc.relation.ispartofInternational Journal of Oral & Maxillofacial Implants
dc.relation.ispartofjcr1.699
dc.relation.ispartofsjr1,576
dc.rights.accessRightsAcesso restritopt
dc.sourceWeb of Science
dc.subjecttorqueen
dc.subjectscrewen
dc.subjectabutmenten
dc.subjectpreloaden
dc.subjectzirconiaen
dc.subjectcyclic loadingen
dc.titleInfluence of Three Types of Abutments on Preload Values Before and After Cyclic Loading with Structural Analysis by Scanning Electron Microscopyen
dc.typeArtigopt
dcterms.rightsHolderQuintessence Publishing Co Inc
dspace.entity.typePublication
relation.isDepartmentOfPublication3936e2e2-946a-42ab-8b9d-9521513200fc
relation.isDepartmentOfPublication.latestForDiscovery3936e2e2-946a-42ab-8b9d-9521513200fc
relation.isOrgUnitOfPublicationca4c0298-cd82-48ee-a9c8-c97704bac2b0
relation.isOrgUnitOfPublication.latestForDiscoveryca4c0298-cd82-48ee-a9c8-c97704bac2b0
unesp.author.lattes6380225075834021
unesp.author.orcid0000-0002-7505-8667[2]
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Odontologia, Araraquarapt
unesp.departmentMateriais Odontológicos e Prótese - FOARpt

Arquivos