Ergonomic Design of Diagnostic Ultrasound Transducers: Analysis, Redesign/Modeling, Simulation/Assessment and Product
Loading...
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Crc Press-taylor & Francis Group
Type
Article
Access right
Abstract
The development of new technology has resulted in more precise medical ultrasound diagnostics; however, the intensive use of poorly designed equipment has been closely linked to musculoskeletal problems among physicians working with ultrasound. This study presents procedures for the analysis, redesign, modeling and simulation/assessment of diagnostic ultrasound transducers. The assessment was performed in four phases: 1 - evaluation during simulated activity and analysis of movement (electrogoniometry) with two commercial transducers; 2 - redesign of the product, in which the aspects of concept product, detailed product and modeling of the mock-ups were considered; 3 - evaluation during simulated activity with one mock-up and with two commercially available transducers. 4 project revision and the development new mock-ups and a subjective assessment (perception scale) of all tested equipment. The project revision resulted in two new mock-ups that required smaller range of motion (P <= 0.05), better load distribution and achieved better acceptability. Systematic methodology of ergonomic design can be used to develop safer and more comfortable products.
Description
Keywords
Language
English
Citation
Advances in Cognitive Ergonomics. Boca Raton: Crc Press-taylor & Francis Group, p. 307-316, 2010.