Development of a pressure nociceptive threshold testing device for evaluation of analgesics in cats

Carregando...
Imagem de Miniatura

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

W B Saunders Co Ltd

Tipo

Artigo

Direito de acesso

Acesso restrito

Resumo

A pressure analgesiometric device was developed for unrestrained cats. Eleven cats were studied. Stimulation was via three rounded pins within a bracelet on the forearm. The pins were advanced by manual bladder inflation. Bladder pressure was measured using a strain gauge pressure transducer. The threshold was recorded at the behavioural end point. Thresholds were measured at 5 and 15 min intervals for 2-4 h, after removal/replacement of the cuff, for 120 min after SC butorphanol (0.4 mg/kg), and with mild skin inflammation at the testing site. Data were analysed using ANOVA. Pressure thresholds in untreated cats were around 150 mmHg. The minimum interval for testing was established as 15 min. Data were reproducible over 4 h and beyond 24 h. Thresholds in 5 cats increased (P < 0.05) above baseline for 45 min after butorphanol with a maximum increase of 270 +/- 182 mmHg at 10 min. Thresholds decreased with inflammation. The method appears suitable for feline analgesia investigations. (c) 2006 Elsevier Ltd. All rights reserved.

Descrição

Palavras-chave

pressure threshold, cat, analgesia, pain, equipment, measurement, pressure nociceptive threshold testing, opioid, NSAID

Idioma

Inglês

Como citar

Research In Veterinary Science. London: W B Saunders Co Ltd, v. 82, n. 1, p. 85-92, 2007.

Itens relacionados

Financiadores

Unidades

Departamentos

Cursos de graduação

Programas de pós-graduação