Electroanalytical determination of carbendazim and fenamiphos in natural waters using a diamond electrode

Nenhuma Miniatura disponível

Data

2012-07-01

Autores

Franca, Rafaela F.
de Oliveira, Hueder Paulo M.
Pedrosa, Valber A. [UNESP]
Codognoto, Lucia

Título da Revista

ISSN da Revista

Título de Volume

Editor

Elsevier B.V. Sa

Resumo

In this study, a method for electroanalytical determination of carbendazim (CBZ) and fenamiphos (FNP) in natural and spiked water was developed using square wave voltammetly in Na2HPO4 0.1 mol L-1 as supporting electrolyte. The calibration curve for carbendazim detection presented good linearity in the concentration range of 0.50 to 15.0 mu M, with a sensitivity of 0.080 A/mol L-1 and a linearity of 0.998. The oxidation of fenamiphos on BDD electrode shows a dynamic range of concentration of 0.5 to 25.0 mu M and sensibility of 0.14 A/mol L-1. The recovery experiments showed values between 70 and 100% for spiked samples thus indicating the feasibility of the electroanalytical methodology to quantify CBZ and FNP in pure or natural waters. (C) 2012 Elsevier B.V. All rights reserved.

Descrição

Palavras-chave

Pesticide, Diamond electrode, Carbendazim, Fenamiphos, Water analysis

Como citar

Diamond and Related Materials. Lausanne: Elsevier B.V. Sa, v. 27-28, p. 54-59, 2012.