Optimization of an amperometric biosensor for the detection of hepatitis C virus using fractional factorial designs

dc.contributor.authorUliana, Carolina V. [UNESP]
dc.contributor.authorRiccardi, Carla S. [UNESP]
dc.contributor.authorTognolli, Joao O. [UNESP]
dc.contributor.authorYamanaka, Hideko [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2014-05-20T14:19:21Z
dc.date.available2014-05-20T14:19:21Z
dc.date.issued2008-01-01
dc.description.abstractFractional factorial design and factorial with center point design were applied to the development of an amperometric biosensor for the detection of the hepatitis C virus. Biomolecules were immobilized by adsorption on graphite electrodes modified with siloxane-poly(propyleneoxide) hybrid matrix prepared using the sol-gel method. Several parameters were optimized, such as the streptavidin concentration at 0.01 mg mL(-1) and 1.0% bovine serum albumin, the incubation time of the electrodes in the complementary DNA solution for 30 minutes and a 1: 1500 dilution of the avidin-peroxidase conjugate, among others. The application of chemometric studies has been efficient, since the best conditions have been established with a restricted number of experiments, indicating the influence of different factors on the system.en
dc.description.affiliationUniv Estadual Paulista, Inst Quim, BR-14800900 Araraquara, SP, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, Inst Quim, BR-14800900 Araraquara, SP, Brazil
dc.format.extent782-787
dc.identifierhttp://dx.doi.org/10.1590/S0103-50532008000400024
dc.identifier.citationJournal of The Brazilian Chemical Society. São Paulo: Soc Brasileira Quimica, v. 19, n. 4, p. 782-787, 2008.
dc.identifier.doi10.1590/S0103-50532008000400024
dc.identifier.fileS0103-50532008000400024.pdf
dc.identifier.issn0103-5053
dc.identifier.lattes1923726000036625
dc.identifier.scieloS0103-50532008000400024
dc.identifier.urihttp://hdl.handle.net/11449/25839
dc.identifier.wosWOS:000256957100023
dc.language.isoeng
dc.publisherSoc Brasileira Quimica
dc.relation.ispartofJournal of the Brazilian Chemical Society
dc.relation.ispartofjcr1.444
dc.relation.ispartofsjr0,357
dc.rights.accessRightsAcesso aberto
dc.sourceWeb of Science
dc.subjectbiosensoren
dc.subjecthepatitis C virusen
dc.subjectfractional factorial designen
dc.titleOptimization of an amperometric biosensor for the detection of hepatitis C virus using fractional factorial designsen
dc.typeArtigo
dcterms.rightsHolderSoc Brasileira Quimica
unesp.author.lattes1923726000036625
unesp.author.lattes0173401604473200[2]
unesp.author.orcid0000-0003-2192-5312[2]
unesp.campusUniversidade Estadual Paulista (Unesp), Instituto de Química, Araraquarapt
unesp.departmentQuímica Analítica - IQARpt

Arquivos

Licença do Pacote

Agora exibindo 1 - 2 de 2
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: