Influence of microwave sterilization on the cutting capacity of carbide burs

Carregando...
Imagem de Miniatura

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Universidade de São Paulo (USP), Faculdade de Odontologia de Bauru

Tipo

Artigo

Direito de acesso

Acesso abertoAcesso Aberto

Resumo

Objective: This study compared the cutting capacity of carbide burs sterilized with microwaves and traditional sterilization methods. Material and Methods: Sixty burs were divided into 5 groups according to the sterilization methods: dry heat (G1), autoclave (G2), microwave irradiation (G3), glutaraldehyde (G4) or control - no sterilization (G5). The burs were used to cut glass plates in a cutting machine set for twelve 2.5-min periods and, after each period, they were sterilized (except G5) following the protocol established for each group. The cutting capacity of the burs was determined by a weight-loss method. Data were analyzed statistically by Kruskal-Wallis and Dunn's test. Results: The means of the cutting amount performed by each group after the 12 periods were G1 = 0.2167 +/- 0.0627 g; G2 = 0.2077 +/- 0.0231 g; G3 = 0.1980 +/- 0.0326 g; G4 = 0.1203 +/- 0.0459 g; G5 = 0.2642 +/- 0.0359 g. There were statistically significant differences among the groups (p<0.05); only dry heat sterilization was similar to the control. Conclusion: Sterilization by dry heat was the method that least affected the cutting capacity of the carbide burs and microwave sterilization was not better than traditional sterilization methods.

Descrição

Palavras-chave

Sterilization, Dental instruments, Infection control, Microwaves

Idioma

Inglês

Como citar

Journal of Applied Oral Science. Bauru-sp: Univ São Paulo Fac Odontologia Bauru, v. 17, n. 6, p. 584-589, 2009.

Itens relacionados

Unidades

Departamentos

Cursos de graduação

Programas de pós-graduação