Progressive X-ray tube degradation detected via daily mammography quality control

Carregando...
Imagem de Miniatura

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Taylor & Francis

Tipo

Artigo

Direito de acesso

Acesso restrito

Resumo

This study evaluated whether routine daily quality control (QC) tests can anticipate X-ray tube degradation in a digital mammography system. Over 30 months, tube loading, contrast-to-noise ratio (CNR), and image quality scores were extracted from daily QC tests and analysed alongside a documented tube failure in September 2024. These results were compared with annual tube output measurements and environmental data. A progressive rise in tube loading was observed prior to tube failure, increasing from 84.45 to 94.58 mAs (12.0%). The Mann-Kendall test confirmed a significant upward trend (τ = 0.461; <i>p</i> &lt; 0.001), with Sen's slope indicating +0.19 mAs/month. Despite this, CNR values remained stable (mean = 14.76 ± 0.20), and image quality showed no degradation. However, tube output declined consistently (from 30.95 ± 3.08 to 28.76 ± 0.36 μGy/mAs), confirming reduced efficiency. Temperature and humidity remained within recommended limits. These findings suggest that gradual increases in tube loading can indicate early tube degradation, even when image quality remains unaffected. Beyond confirming equipment performance, daily QC may serve as a strategic tool to detect early signs of degradation, extend device lifespan, reduce unplanned downtime, and promote patient safety without added costs.

Descrição

Palavras-chave

Citação

Itens relacionados

Financiadores

Unidades

Tipo de item:Unidade,
Botucatu, Faculdade de Medicina - FMB
FMB
Campus: Botucatu

Tipo de item:Unidade,
Botucatu, Instituto de Biociências - IBB
IBB
Campus: Botucatu

Departamentos

Cursos de graduação

Programas de pós-graduação

Outras formas de acesso