Logotipo do repositório

Progressive X-ray tube degradation detected via daily mammography quality control

dc.contributor.authorMarques, Túlio Guilherme Soares [UNESP]
dc.contributor.authorde Pina, Diana Rodrigues [UNESP]
dc.contributor.authorAlvarez, Matheus [UNESP]
dc.date.accessioned2026-06-18T19:11:52Z
dc.date.issued2025-10-07
dc.description.abstractThis study evaluated whether routine daily quality control (QC) tests can anticipate X-ray tube degradation in a digital mammography system. Over 30 months, tube loading, contrast-to-noise ratio (CNR), and image quality scores were extracted from daily QC tests and analysed alongside a documented tube failure in September 2024. These results were compared with annual tube output measurements and environmental data. A progressive rise in tube loading was observed prior to tube failure, increasing from 84.45 to 94.58 mAs (12.0%). The Mann-Kendall test confirmed a significant upward trend (τ = 0.461; <i>p</i> &lt; 0.001), with Sen's slope indicating +0.19 mAs/month. Despite this, CNR values remained stable (mean = 14.76 ± 0.20), and image quality showed no degradation. However, tube output declined consistently (from 30.95 ± 3.08 to 28.76 ± 0.36 μGy/mAs), confirming reduced efficiency. Temperature and humidity remained within recommended limits. These findings suggest that gradual increases in tube loading can indicate early tube degradation, even when image quality remains unaffected. Beyond confirming equipment performance, daily QC may serve as a strategic tool to detect early signs of degradation, extend device lifespan, reduce unplanned downtime, and promote patient safety without added costs.
dc.description.affiliationInstitute of Biosciences, São Paulo State University (UNESP), Botucatu, São Paulo, Brazil
dc.description.affiliationBotucatu Medical School Clinics Hospital, São Paulo State University (UNESP), Botucatu, Brazil
dc.description.affiliationDepartment of Infectious Diseases, Dermatology, Imaging Diagnosis, and Radiotherapy, São Paulo State University (UNESP), Botucatu, Brazil
dc.description.affiliationUnespInstitute of Biosciences, São Paulo State University (UNESP), Botucatu, São Paulo, Brazil
dc.description.affiliationUnespBotucatu Medical School Clinics Hospital, São Paulo State University (UNESP), Botucatu, Brazil
dc.description.affiliationUnespDepartment of Infectious Diseases, Dermatology, Imaging Diagnosis, and Radiotherapy, São Paulo State University (UNESP), Botucatu, Brazil
dc.identifierhttps://app.dimensions.ai/details/publication/pub.1193644212
dc.identifier.dimensionspub.1193644212
dc.identifier.doi10.1080/03091902.2025.2553132
dc.identifier.issn0309-1902
dc.identifier.issn1464-522X
dc.identifier.orcid0000-0002-6852-2724
dc.identifier.orcid0000-0003-1967-1990
dc.identifier.orcid0000-0003-4775-0194
dc.identifier.pmid41055117
dc.identifier.urihttps://hdl.handle.net/11449/326240
dc.publisherTaylor & Francis
dc.relation.ispartofJournal of Medical Engineering & Technology; n. ahead-of-print; v. ahead-of-print; p. 1-8
dc.rights.accessRightsAcesso restritopt
dc.rights.sourceRightsclosed
dc.sourceDimensions
dc.titleProgressive X-ray tube degradation detected via daily mammography quality control
dc.typeArtigopt
dspace.entity.typePublication
relation.isOrgUnitOfPublicationa3cdb24b-db92-40d9-b3af-2eacecf9f2ba
relation.isOrgUnitOfPublicationab63624f-c491-4ac7-bd2c-767f17ac838d
relation.isOrgUnitOfPublication.latestForDiscoverya3cdb24b-db92-40d9-b3af-2eacecf9f2ba
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Biociências, Botucatupt
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Medicina, Botucatu

Arquivos